Walk in Chamber for Rapid Rate Temperature / Humidity Test

The Walk in Chamber for Rapid Rate Temperature / Humidity Test is applicable to the standard  UL 2580.

Technical Parameters

Product Name Rapid temperature change rate (damp-heat) Test Chamber
Model BTKS5-12m³-C
Internal Dimension W2000 X D3000 X H2000mm
External Dimension W2400 X D4500 X H2940mm
Temperature Range -50℃ ~ +150℃ (Adjustable)
Humidity Range 20%RH-98%RH (Limit temperature section, see the regional figure )Rapid Temperature Change Rate Humidity Temp. Test Chamber3 300x146.jpg
Temperature Fluctuation ±0.5℃
Humidity Fluctuation ±2.0%RH
Temperature Deviation 2.0℃
Humidity Deviation A) >75%RH: ≤+2, -3%RH, B) <75%RH: ≤±5%RH.
Heating Rate ≧5℃/min (From -40℃ up to +80℃, No load)
Cooling Rate ≧5℃/min (From -40℃ up to +80℃, No load)
Temperature overshoot ≤±2℃
Weight Approx 5 Tons
Power Supply AC 380V, 50HZ, 127KW
Noise ≤70dB (A-Level)
External Material Galvanized sheet with high temperature electrostatic resin coating at both sides.
Internal Material 1.2 mm thickness SUS# 304 heat and cold resistant stainless steel with seal welding.
Controller 24-bit high accuracy 7 inch TFT true colors LCD touch screen.Rapid Temperature Change Rate Humidity Temp. Test Chamber4.png
Compressor Semi-Hermetic Compressors BOCK / BITZER (Germany)
Cooling Type Air cooling or water cooling (Optional)
Other Requirement Main-subsidiary type trolley for the specimen in and out of the test chamber.Explosion-proof lighting system.

Explosion-proof pressure relief system.

Smoke exhaust system.

Automatic Water Supply System.

Fire extinguishing device; spray device; smoke alarm device;

 

Temperature shock

  1. Test Object: Battery storage pack or system.
  2. Li-ion battery pack or system placed in (-40±2) ℃~ (85±2) ℃ alternation temperature, the conversion time of two extreme temperatures is less than 30 minutes. Place the test objects into each extreme temperature keeps 8h, test 5 cycles.
  3. Observe for 2 hours under the room temperature after testing,

 

Humidity & Temperature Cycling

  1. Test Object: Battery storage pack or system.
  2. Refer to the GB/T2423.4 Test Db, variable refer the Graph 4, the highest temperature is +80℃, 5 cycles. Observe for 2h under room temperature.

 

Rapid Temperature Change Rate Humidity Temp. Test Chamber1 300x204.png Rapid Temperature Change Rate Humidity Temp. Test Chamber2 300x183.png

Y1: Relative humidity %RH;    Y2: Temperature, ℃;       X: Time, h;

a: end of the heating;      b: cooling start;

c: recommended temperature and humidity values;

d: condensation;         e: desiccation;        f: 1 cycle.

 

Standard

  • UL 2580