Thermal Abuse Test Chamber – Test Battery Module

Bellue’s BE-101-1000A chamber supports controlled thermal abuse testing of battery modules. Define the specimen, loaded heating profile, monitoring, protective configuration, and applicable test procedure when requesting a quotation.

Thermal Abuse Test Chamber Technical Parameters

Model BE-101-1000A
Temp. Range Ambient – 200℃ (Controllable)
Control Accuracy ±0.5℃
Temp. Uniformity ±2.0℃( No load)
Temp. Rising Speed RT – 150℃ (≧5℃/min, liner with load 40kg module)
Temp. overshoot ≦2.0℃
Control Display Method PLC Touch screen + remote control, networking,
Cell phone APP alarm device, with USB interface, R485 interface can be connected to computer control.
Testing Room Dimension W1000 X D1000 X H1000 mm
External Chamber Dimension W1400 X D1300 X H1840 mm
Inner chamber Material SUS304 Stainless Steel (Mirror finished)
External chamber Material SECC Steel Plate, Powder Painted
Basement With universal wheel
Observation Window 390 X 360mm (20mm Tempered Explosion Proof Glass)
Power Source 380V 50HZ
Heating Power About 12 KW
Safety Protection Device over-voltage, over-temperature, over-current protection.
protection for short circuits caused by incorrect operation of the operator and battery electrode, battery positive and negative connection, etc.
Safety Device Smoke exhaust system, explosion-proof pressure relief port, tri-color light alarm.

 

Thermal Abuse Test Chamber Standards

  • IEC 62133
  • IEC 62619
  • UL 1642

 

Define the Cell or Module Test Program

Confirm the battery chemistry, application, specimen level, applicable standard edition, and approved procedure before defining the exposure program. IEC 62133-2 addresses portable lithium systems, while IEC 62619 addresses industrial lithium cells and batteries; the required method must match the project scope.

  • Specimen dimensions, mass, quantity, state of charge, and preparation requirements.
  • Required chamber temperature profile, loaded heating rate, dwell time, and measurement locations.
  • Voltage and temperature monitoring, observation, data recording, and acceptance criteria.
  • Fixture arrangement, feedthroughs, exhaust interfaces, and protective equipment configuration.
  • Laboratory-approved procedure, installation requirements, and operator training scope.

Request engineering review of the complete test program. The chamber parameter table describes equipment configuration; the selected procedure defines preparation, exposure, observation, and evaluation requirements.