Thermal Abuse Test Chamber – Test Battery Module

Thermal abuse test chamber protect for short circuit caused by incorrect operation of the operator and battery electrode, battery positive and negative connection, etc

Thermal Abuse Test Chamber Technical Parameters

Model BE-101-1000A
Temp. Range Ambient – 200℃ (Controllable)
Control Accuracy ±0.5℃
Temp. Uniformity ±2.0℃( No load)
Temp. Rising Speed RT – 150℃ (≧5℃/min, liner with load 40kg module)
Temp. overshoot ≦2.0℃
Control Display Method PLC Touch screen + remote control, networking,
Cell phone APP alarm device, with USB interface, R485 interface can be connected to computer control.
Testing Room Dimension W1000 X D1000 X H1000 mm
External Chamber Dimension W1400 X D1300 X H1840 mm
Inner chamber Material SUS304 Stainless Steel (Mirror finished)
External chamber Material SECC Steel Plate, Powder Painted
Basement With universal wheel
Observation Window 390 X 360mm (20mm Tempered Explosion Proof Glass)
Power Source 380V 50HZ
Heating Power About 12 KW
Safety Protection Device over-voltage, over-temperature, over-current protection.
protection for short circuits caused by incorrect operation of the operator and battery electrode, battery positive and negative connection, etc.
Safety Device Smoke exhaust system, explosion-proof pressure relief port, tri-color light alarm.

 

Thermal Abuse Test Chamber Standards

  • IEC 62133
  • IEC 62619
  • UL 1642

 

Single storage cell Safety Test

Heating test steps:

  • Charge the battery according to the standard 6.1.3
  • Place it into the chamber

–For lithium storage battery, the oven temperature is raised at a rate of 5℃/min to the temperature of 130℃±2℃, remain this temperature for 30min, then stop heating.

–For metal cyanide nickel battery, the oven temperature is raised at a rate of 5℃/min to a temperature of 85℃±2℃, remain this temperature for 2h, then stop heating.

  • Observe for 1h

 

Storage Battery Module Safety Test

  • Charge the battery module according to the standard 6.1.4
  • Place it into the chamber

-For lithium storage battery, the oven temperature is raised at a rate of 5℃/min to a temperature of 130℃±2℃, remain this temperature for 30min, then stop heating.

–For metal cyanide nickel battery, the oven temperature is raised at a rate of 5℃/min to a temperature of 85℃±2℃, remain this temperature for 2h, then stop heating.

  • Observe for 1h